Manual Mode - Memmert UN PLUS Series Operating Instructions Manual

Hide thumbs Also See for UN PLUS Series:
Table of Contents

Advertisement

The status display shows you which operating mode or operating state the appliance is
currently in. The current operating state is highlighted in colour and indicated by the text
display:
Appliance is in programme mode
Programme is stopped
Appliance is in manual operating mode
The example on the right shows the appliance in man-
ual mode, identified by the coloured hand symbol.

5.4.4 Manual mode

In this operating mode, the appliance runs in permanent operation at the values set on the
ControlCOCKPIT.
Adjustment options
As described in chapter 5.4.2 , you can set the following parameters after pressing the cor-
responding activation key (in any sequence):
Temperature
Adjustment range: model dependent (see nameplate and technical
date on page 14)
Heating operation is indicated by the
You can select °C or °F as the temperature unit displayed (see
page 40).
Air flap position
Adjustment range: 0 % (closed, recirculating operation) to 100 %
(completely opened, fresh air operation) in steps of 10 %
Fan speed
(only for UF/IF/SF
PLUS
Adjustment range: 0 to 100 % in steps of 10%
Interior lighting (additional option)
Adjustment range: 0 %, 100 %
appliances)
12.Sept.2012
symbol.
Operation and control
13:44
Manual Mode
TEMP
24
180
Set
FLAP
40
F
FA
FAN
N
50
LIGHT
100
.4 °C
.4
°C
%
%
%
25

Advertisement

Table of Contents
loading