5.4.6 Programme mode

In this operating mode, programmes saved in the appliance can be started with different
combinations of individual parameters (temperature, air flap position, fan speed, working
chamber lighting) at staggered intervals, which the appliance then automatically processes in
sequence. These programmes are not created directly at the appliance but externally at a com-
puter / laptop and using AtmoCONTROL software. Transfer to the appliance is possible using
the provided USB storage medium or via Ethernet.
A description of how to create and save programmes can be found in the separate
AtmoCONTROL software manual.
Starting a programme
1. Press the activation key to the right of
the status display. The current operating
mode is highlighted automatically, in this
Manual Mode
example
2. Turn the turn control until the
symbol is highlighted. The current
programme is displayed, in this example
Test 012
.
Only the programme currently selected in the menu and shown in the display can be
used. If you want to process another programme, you need to activate it in the menu
first (description from page 48 ).
3. To start the programme, press the confir-
mation key. The programme starts. The
display shows:
the programme description (in this exam-
Test 012
ple
)
the programme segment description, in
Ramp 1
this example
the current run (in case of loops)
You cannot change any parameters (e.g. the temperature) at the appliance while a
programme is running. However, the displays
Cancelling a programme
You can cancel an active programme at any
time.
1. Press the activation key to the right of
the status display. The status display is
automatically highlighted.
0
(
).
0°C
off
start
0
0°C
Fr 20.10.2010 20:31
13.Sept.2012
17:44
Manual mode
manueller Betrieb
%
%
GRAPH
GRAPH
APH
80
80
40
40
ALARM
GRAPH
and
Fr 20.10.2010 20:31
12.Sept.2012
10:44
manueller Betrieb
Test 012
%
%
Ramp 3
GRAPH
GRAPH
APH
80
80
Operation and control
12.Sept.2012
10:44
Test 012
ready
12.Sept.2012
10:44
Test 012
Ramp 1
can still be used.
27

Advertisement

Table of Contents
loading

Table of Contents