Operation and control
2. By turning the turn control, adjust the
desired lower alarm limit value, in the
example on the right 160 °C.
If no undertemperature protection limit is
required, set the lowest temperature.
3. Press the confirmation key to confirm.
max
The
display (overtemperature pro-
tection) is activated.
4. By turning the turn control, adjust the
desired upper alarm limit value, in the
example on the right 190 °C.
The monitoring temperature must be set
sufficiently high above the maximum set
temperature. We recommend 5 to 10 K.
5. Accept the upper alarm limit value by
pressing the confirmation key. The setting
of the automatic temperature monitor
(ASF) is automatically activated (
6. With the turn control, select ON (
OFF ().
7. Press the confirmation key to confirm.
The ASF tolerance band setting is acti-
vated.
8. With the turn control, adjust the desired
tolerance band, e.g. 5.0 K.
We recommend a tolerance band of 5 to
10 K (1 to 3 K for incubators IN/IF).
9. Press the confirmation key to confirm.
Temperature monitoring is now active.
32
auto
).
) or
ALARM
min
max
160
120
0 .
°C
.0 °C
auto
+ / - 0.0
ALARM
min
max
160
120
0 .
°C
0 .
auto
+ / - 0.0
ALARM
min
max
160
190
0 .
°C
0 .
auto
+ / - 0.0
ALARM
min
max
160
190
.0 °C
.0 °C
auto
+ / - 0.0
ALARM
min
max
160
190
.0 °C
.0 °C
auto
+ / - 0.0
ALARM
min
max
160
190
0 .
°C
.0 °C
auto
+ / -
0.0
ALARM
min
max
160
190
0 .
°C
.0 °C
auto
+ / -
5.0
ALARM
min
max
160
190
.0 °C
.0 °C
auto
+ / - 5.0
K
°C
K
°C
K
K
K
K
K
K

Advertisement

Table of Contents
loading

Table of Contents