4. Save the setting by pressing the
confirmation key.

7.3.6 Balance

Description
For appliances of the size 55 and above, application-specific correction of the heat output
distribution (balance) between the upper and lower heating groups is possible. The
adjustment range is from –50 % to +50 %.
lower heat output
IP address
Subnetmask
Unit
Alarm temp
Timer Mode
Balance
Fig. 24
Heat output distribution (example): The –20 % (left) setting causes the lower heating
groups to emit 20 % less heat than the upper ones. The +30 % (right) setting causes the lower
heating groups to emit 30 % more heat than the upper ones. The 0 % setting restores the
default heat output distribution.
Setting
SETUP
1. Activate the
Balance
with the turn control.
upper heat output
-20%
SETUP
192. 168. 1 0 0 . 100
255. 255. 0 . 0
°C
F
TWW
TWB
-20
%
display and select
IP address
Subnetmask
Unit
Alarm temp
Timer Mode
Balance
upper heat output
lower heat output
+30%
SETUP
IP address
192. 168. 1 0 0 . 100
Subnetmask
255. 255. 0 . 0
Unit
°C
Alarm temp
TWW
Timer Mode
Balance
+30
%
IP address
Subnetmask
Unit
Alarm temp
Timer Mode
Balance
Menu functions
255. 145. 1 3 6 . 225
255. 255. 0 . 0
°C
F
TWW
TWB
-20 %
F
TWB
192. 168. 1 0 0 . 100
255. 255. 0 . 0
°C
F
TWW
TWB
+30
%
43