Vt Thermal Exchangeable Mount - NT-MDT NTEGRA Therma Instruction Manual

Probe nanolaboratory configuration with variable temperature assembly for thermal measurements
Table of Contents

Advertisement

NTEGRA Therma Probe NanoLaboratory. Instruction Manual
The desing of Ntegra Therma main units is discussed in the books «Performing
Measurements» and «Control Electronics». The features of configuration for wide-range
temperature measurements are described below.
Specifications
Parameter
Temperature range
Stability of temperature maintenance
Thermal drift on XY
Thermal drift on Z
Scan range
XY linearity error
Z linearity error
RMS X, Y sensor noise in 300 Hz band
RMS Z sensor noise in 1 kHz band
RMS noise on Z in 1 kHz band
Maximum sample size

1.1. VT thermal Exchangeable Mount

A general view of the VT thermal exchangeable mount is given in Fig. 1-2. The
exchangeable mount provides both low temporal drift in the temperature range from -30°С
to +80°С and small temperature drift after the temperature of the cell has been changed.
1 – temperature chamber; 2 –heat exchanger of scanner; 3 – outlets for liquid cooling
6
Fig. 1-2. VT thermal exchangeable mount
system; 4 – registration head seats
Value
from -30°C up to +80°C
0.005°C typically
better than 0.001°C
≤15 nm/°C
≤10 nm/°C
10x10x6 μm
0.2% peak-to-peak/2
1% peak-to-peak/2
≤0.03 nm
≤0.05 nm
<0.025 nm
∅14x1 mm