Tec Calibration; Recommended Equipment; Local Operation Thermistor Calibration; Table 11 - Recommended Test Equipment - Newport 6000 Operation And Maintenance Manual

Laser controller
Table of Contents

Advertisement

80
Chapter 7
7.3

TEC Calibration

This chapter describes how to calibrate the TEC modules.
7.3.1

Recommended Equipment

Recommended test equipment for calibrating the module is listed in Table 11.
Equipment other than that shown in the table may be used if the specifications meet
or exceed those listed.
Description
DMM
Resistors
Resistor
Connector
7.3.2

Local Operation Thermistor Calibration

Calibration
h.
Once the actual voltage value is entered via the LAS:LDV command,
the 6000 leaves the current calibration mode.
If, at any time prior to the second LAS:LDV, a command other than
LAS:LDV or LAS:CAL:LDV? is sent to the 6000, the 6000 will
cancel the calibration mode and then process the command(s).
The *OPC? query may be used to determine when the calibration is
completed.
The operation complete flag (bit 0 of the Standard Event Status
Register) may be used to trigger an interrupt. This type of interrupt is
enabled by setting bit 0 of the Service Request Enable register and
using the *OPC command.
Mfg./Model
HP34401A
Metal Film
High Power
D-sub

Table 11 - Recommended Test Equipment

a.
Measure and record the exact resistance of your metal film resistor. Use
nominal values of 20 kΩ for the 100µA setting, and 200 kΩ for the
Specification
DC Amps @ 1.0 A): ±1%
Resistance (@ 10 ohms): 0.02%
20 kΩ for 100µA calibration
200 kΩ for 10µA calibration
3 kΩ for LM335 sensor calibration
16 kΩ for AD590 sensor calibration
100 Ω for RTD sensor calibration
1 Ω, 50 W, for current calibration
15-pin male