Olympus OEP-4 Instructions Manual page 68

Color video printer
Hide thumbs Also See for OEP-4:
Table of Contents

Advertisement

Irregularity
The number displayed on
the CVP counter does not
increase even if an image
has been captured.
Images cannot be printed.
7
The HDTV image with the
16:9 aspect ratio is printed
improperly.
The color tones in the print
picture are incorrect.
68
7.1 Troubleshooting Guide
Possible cause
The communication problem has occurred
between this instrument and the connected
equipment.
The connected equipment is set improperly.
The communication speed (BAUDRATE) is set
improperly.
The remote cable is connected improperly.
The remote cable is broken.
The remote cable is subjected to an excessive
force.
The endoscope or the scope cable is connected
improperly.
The video cable is connected improperly.
The video cable is broken.
The video cable is subjected to an excessive
force.
No image has been captured.
No ink ribbon is installed.
The ink ribbon is installed improperly.
This ink ribbon has been torn.
No print sheets are installed.
The print sheets are installed improperly.
"AUTO PRINT" is set to "OFF."
The input signal is switched.
Only the "4 : 3" and "5 : 4" aspect ratios are
available.
Some setups are incorrect.
Solution
Turn this instrument OFF, then ON again.
Turn the connected equipment OFF, then ON again.
Set the connected equipment properly as described in
"3.1 Setup of Connected Equipment" on page 25, or in
"3.3 Setup of Connected Equipment" in
"INSTRUCTIONS (BASIC)."
Set it properly as described in "Setting "BAUDRATE""
on page 59 and in the instruction manual of the connected
equipment.
Connect the remote cable properly as described in
"Chapter 2 Installation and Connection" on page 15 or in
"Chapter 2 Installation and Connection" in
"INSTRUCTION (BASIC)."
Replace the new remote cable.
Change the cable layout to remove the excessive force.
Connect the endoscope or the scope cable properly as
described in the instruction manual of the connected
equipment.
Connect it properly as described in "Chapter 2
Installation and Connection" in "INSTRUCTIONS
(BASIC)" or in "INSTRUCTIONS (ADVANCED)."
Replace the new video cable.
Change the cable layout to remove the excessive force.
Capture an image.
Install the ink ribbon as described in "Installation of the
Ink Ribbon" in "3.2 Installation of the Color Printing
Pack (UPC-55)" in "INSTRUCTIONS (BASIC)."
Install the ink ribbon as described in "Installation of the
Ink Ribbon" in "3.2 Installation of the Color Printing
Pack (UPC-55)" in "INSTRUCTIONS (BASIC)."
Repair the tear as described in "If the ink ribbon is torn in
the middle" in "3.2 Installation of the Color Printing Pack
(UPC-55)" in "INSTRUCTIONS (BASIC)."
Install the ink ribbon as described in "Installation of the
Print Sheets" in "3.2 Installation of the Color Printing
Pack (UPC-55)" in "INSTRUCTIONS (BASIC)."
Install the ink ribbon as described in "Installation of the
Print Sheets" in "3.2 Installation of the Color Printing
Pack (UPC-55)" in "INSTRUCTIONS (BASIC)."
Set "AUTO PRINT" to "ON" or "ON CLEAR," referring
to "Setting "AUTO PRINT"" on page 43.
Perform printing manually from the front panel of this
instrument or by a remote operation from the connected
equipment.
Set the signal format of the switched input signal,
referring to "5.8 Setup of the INPUT SETUP Menu" on
page 49.
This is the design of this instrument and not a
malfunction.
Adjust the color tones as described in the description on
"5.10 Setup of the COLOR BALANCE Menu" on page
52 in "INSTRUCTIONS (ADVANCED)."