PROCEdURE 4.
CHARGE THE MONITOR BATTERY
WARNING
In order to maintain electrical safety, use only the provided, medical‑approved power supply.
The Cobalt AVL Monitor includes an internal lithium‑ion battery. Verathon
battery fully prior to first use.
Under normal operating conditions, a fully charged battery lasts approximately 90 minutes or longer before it
needs to be recharged. For optimal battery life, ensure that the battery is fully charged before you try to use
the monitor in battery mode. You should charge the battery at temperatures between 0–35°C (32–95°F).
Figure 11.
Battery Status Icons
Battery must be charged
Approximately ⅓ battery life remains
Approximately ⅔ battery life remains
Battery is ⅔ to fully charged
1. Connect the video monitor 12 V DC power adapter to the power cord.
2. On the back panel of the monitor, remove the power socket cap, and then connect the 12 V DC power
adapter to the power socket.
3. Plug the power supply into a hospital‑grade power outlet.
4. Allow the battery to charge. Fully charging the battery may take up to 6 hours.