PROCEdURE 4.
WARNING
In order to maintain electrical safety, use only the provided, medical‑approved power supply.
The GlideScope Video Monitor includes an internal lithium‑ion battery. Verathon
charge the battery fully prior to first use.
Under normal operating conditions, a fully charged battery lasts approximately 90 minutes or longer before it
needs to be recharged. For optimal battery life, ensure that the battery is fully charged before you try to use
the monitor in battery mode. You should charge the battery at temperatures between 0–35°C (32–95°F).
The percentage above the Battery Status icon indicates the remaining battery charge.
Figure 11.
Battery Status Icons
19% battery life or less remaining. Battery must be charged.
20% to 50% battery life remaining.
51% to 82% battery life remaining.
Battery is 83% to fully charged. The lightning bolt indicates that the battery is charging.
1. Connect the video monitor 12 V DC power adapter to the power cable.
2. On the back panel of the monitor, remove the power socket cap, and then connect the 12 V DC power
adapter to the power socket.
3. Plug the power supply into a hospital‑grade power outlet.
4. Allow the battery to charge. Fully charging the battery may take up to 6 hours.
Operations & Maintenance Manual: Setting up
CHARGE THE MONITOR BATTERY
®
recommends that you
21