Electromagnetic Compatibility - Verathon GlideScope Cobalt AVL Single-Use Operation & Maintenance Manual

Table of Contents

Advertisement

ELECTROMAGNETIC COMPATIBILITY

GlideScope AVL system is designed to be in compliance with IEC 60601‑1‑2:2007, which contains
electromagnetic compatibility (EMC) requirements for medical electrical equipment. The limits for emissions
and immunity specified in this standard are designed to provide reasonable protection against harmful
interference in a typical medical installation.
The GlideScope AVL system complies with the applicable essential performance requirements specified in
IEC 60601‑1. Results of immunity testing show that the essential performance of the system is not affected
under the test conditions described in the following tables. For more information about the essential
performance of the GlideScope AVL system, see
ELECTROMAGNETIC EMISSIONS
Table 8.
Guidance and Manufacturer's Declaration—Electromagnetic Emissions
The GlideScope AVL system is intended for use in the electromagnetic environment specified below. The
customer or the user of the GlideScope AVL system should ensure that it is used in such an environment.
EMISSIONS TEST
RF emissions
CISPR 11
Harmonic emissions
IEC 61000‑3‑2
Voltage fluctuations/
flicker emissions
IEC 61000‑3‑3
ELECTROMAGNETIC IMMUNITY
Table 9.
Guidance and Manufacturer's Declaration —Electromagnetic Immunity
The GlideScope AVL system is intended for use in the electromagnetic environment specified below. The
customer or the user of the GlideScope AVL system should ensure that it is used in such an environment.
IMMUNITY TESTS
Electrostatic discharge
(ESD)
IEC 61000‑4‑2
Electrical fast transient/
burst
IEC 61000‑4‑4
Surge
IEC 61000‑4‑5
Operations & Maintenance Manual: Product Specifications
Essential Performance
COMPLIANCE
Class A
The GlideScope AVL system is suitable for use in all establishments
Class A
other than domestic and those directly connected to the public
low‑voltage power supply network that supplies buildings used for
domestic purposes.
Complies
IEC 60601 TEST LEVEL
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
on page 1.
ELECTROMAGNETIC ENVIRONMENT – GUIDANCE
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT
LEVEL
Floors should be wood, concrete,
or ceramic tile. If floors are
In compliance
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
In compliance
that of a typical commercial or
hospital environment.
Mains power quality should be
In compliance
that of a typical commercial or
hospital environment.
– GUIDANCE
47