Movita Lift: Emergency Operation Of The System - Dräger Movita Instructions For Use Manual

Ceiling supply units
Hide thumbs Also See for Movita:
Table of Contents

Advertisement

Fault
Insufficient gas supply, hissing
sound.
Lift system: LED indicator is not lit
green:
Lift system does not respond
when actuated.
Lift system does not respond to
operation of the down arrow key.

Movita lift: Emergency operation of the system

In the event of a power supply failure, the lift system
of the Movita lift version can also be moved without
using the electrically-operated height adjustment.
A
Instructions for use Movita
Cause
Pressurized gas is escaping from
inside the ceiling supply unit.
Board/System without power,
cable connection interrupted.
LED indicator defective.
Defective fuses on the board.
System without power, cable
connection interrupted.
Motor control defective.
Incorrect setting position of the
keyswitch.
Docked medical device and
position of the keyswitch do not
correspond.
Perseus safety sensor triggered. Remove obstacle and operate
1 Take off the cover (A) (see assembly
instructions).
2 Actuate the motor shaft (B) using a 7 mm
socket wrench.
NOTE
The system moves slowly due to the gear ratio.
B
Troubleshooting
Remedy
Notify hospital maintenance or
DrägerService to check the gas
supply. Connected devices must
also be taken into consideration.
Check power supply and fuses
(hospital maintenance).
Replace control panel
(DrägerService).
Replace fuses
(DrägerService).
Check power supply and fuses
(hospital maintenance).
Inform DrägerService.
Contact service personnel.
up arrow key.