Dräger Movita Instructions For Use Manual page 19

Ceiling supply units
Hide thumbs Also See for Movita:
Table of Contents

Advertisement

Connecting the cable remote control
(optional)
WARNING
Risk of electric shock
Electric shock may result if damaged cables
are put in the cable organizer.
Make sure that damaged cables are not put in
the cable organizer.
WARNING
Risk of electric shock
Electric shock may result if single-insulated
cables are put in the cable organizer.
Make sure that only double-insulated cables
are put in the cable organizer.
WARNING
Risk of personal injury and equipment
damage
If too many cables are guided through the
cable organizer, cable damage may occur.
Make sure that the maximum number of
cables and hoses is not exceeded. A
maximum of 5 cables, each with a diameter of
7 mm, and 1 hose, with a diameter of 15 mm,
is allowed.
Instructions for use Movita
C
The cable remote control is available for ceiling
supply units.
1 Plug connector (A) into the receptacle. Make
sure that the key of the connector goes into the
groove of the receptacle.
2 Rotate the knurled screw (B) clockwise =
The connector is locked in place.
 Hang the cable remote control (C) on the rail
(D).
Assembly and preparation
D
B
A
19