User Calibration - Adam Equipment CBD Series Service Manual

Table of Contents

Advertisement

7.0

USER CALIBRATION

ACTION
Press the [Tare/Zero] key during the
self-test at power on. The scale will
ask to enter the password.
Default Password is 0000.
Enter "0" four times.
Password can be changed in technical
parameters.
Press [Print].
Select the scale to be set up by using
the [Local/Remote] key.
Press the [Print] key to enter the
technical section.
Use the [U.Wt.] to select the weighing
unit to be used for setting of the
calibration. The arrow in the "Weight"
window will indicate the unit selected.
Press the [Print] key to continue.
You will enter the first parameter -
Calibration. Display will ask to unload
any weight on the platform.
Press the [Print] key to continue.
Enter the calibration weight to be
loaded and press [Print].
Load the calibration weight onto the
scale and press [Print] again.
The Calibration is complete, the scale
will run the self-test during which the
weight should be removed.
@Adam Equipment Company 2007
DISPLAYS
" Pi n"
" Pi n" " ----"
"tEch" " LocAL" "
"tEch" "rEmo tE" "
"tEch" " Uni t" "
"Un L o Ad"
"SEL" " 0010"
"Lo Ad"
9
"
"
"