Adam Equipment
CBC SERIES
Service Manual
Software rev.: 3.06 & above
ADAM EQUIPMENT CO. LTD.
P.N. 8684, Rev. A2, June 2007
© Adam Equipment Company 2007

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CBC Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Adam Equipment CBC Series

  • Page 1 Adam Equipment CBC SERIES Service Manual Software rev.: 3.06 & above ADAM EQUIPMENT CO. LTD. P.N. 8684, Rev. A2, June 2007 © Adam Equipment Company 2007...
  • Page 2 © Adam Equipment Company 2007...
  • Page 3: Table Of Contents

    6.2.5 F6 -SUCCESSIVE TARE ..................14 6.2.6 F7 -A/D COUNT....................15 6.2.7 F8 –ZERO......................15 6.2.8 F9 –PIN......................15 6.2.9 F10 –LOW VOLTAGE DETECTION ..............16 6.2.10 WIRING DIAGRAM ....................17 MECHANICAL ASSEMBLY ..................18 LIST OF MECHANICAL ASSEMBLY..............19 DISPLAY ASSEMBLY.....................20 ***** © Adam Equipment Company 2007...
  • Page 4 © Adam Equipment Company 2007...
  • Page 5: Introduction

    1.0 INTRODUCTION • The CBC series provides accurate, fast and versatile counting scales. • There are 2 type of scales within the CBC series- CBC scales are gram / kilogram only scales and the CBCa scales are changeable from pounds to kilograms. The scales have the same functions except that CBCa scales have the ability to toggle between the two weighing units.
  • Page 6: Specifications

    Model: CBC 70a UNITS OF MAXIMUM TARE READABILITY REPEATIBILITY LINEARITY MEASURE CAPACITY RANGE Kilograms 32.000 kg -32 kg 0.001 kg 0.001 kg 0.002 kg Pounds 70.000 lb -70 lb 0.002 lb 0.002 lb 0.004 lb © Adam Equipment Company 2007...
  • Page 7: Common Specifications

    Parts counting, check-counting, weighing, accumulating memory, pre-set count with alarm Interface RS-232 bi-directional interface English, German, French, Spanish selectable text Date/Time Real Time Clock (RTC), To print date and time information (Dates in year/month/day, day/month/year or month/day/year formats- Battery backed) © Adam Equipment Company 2007...
  • Page 8: Troubleshooting

    Incorrect adjustment of the mechanical stops. An error “ ” Load cell damage. displays on the screen Scale not installed properly. Incorrect adjustment of the mechanical stops. Factory calibration has been tampered with- check the ADC counts and re-calibrate. © Adam Equipment Company 2007...
  • Page 9: Error Codes

    +10% factory calibration). The old calibration data will be retained until the calibration process is complete. Err 9 Scale is unstable. There is vibration or draft making the scale unstable. Electronics may be damaged. © Adam Equipment Company 2007...
  • Page 10: Cbc Description

    4 strain gage sensors mounted in a particular way on the metal structure. These 4 strain gauges are connected in such a way that they form a wheatstone bridge. See figure below. In the simplest load cells there are only the 4 resistances of the strain © Adam Equipment Company 2007...
  • Page 11 Most load cells used in scales are not as simple as this example. They have additional resistance elements added to compensate for temperature variations and to set the outputs to correct voltages. The circuit of a typical load cell is shown below- © Adam Equipment Company 2007...
  • Page 12 If the load cell has been damaged the no load voltage will likely be greater than 3mv or the loaded voltage will be grossly different from the expected value. Remember the loaded voltage will be offset by the amount of zero load voltage measured. © Adam Equipment Company 2007...
  • Page 13: Service Parameters

    Press the [Zero] key to exit a parameter. Pressing [Zero] again will return the scale to weighing. When a parameter is entered by pressing the [Tare] key, the display will guide you through the parameter selected and the options available. © Adam Equipment Company 2007...
  • Page 14: F1 -Calibration

    FAIL L The display will show “ ” if the calibration is OK. Or it will show “ ” if calibration could not be completed correctly. Remove the calibration weight. The scale will return to weighing. © Adam Equipment Company 2007...
  • Page 15: F2 -Resolution

    35 kg load cells and the 48 kg unit uses 50 kg load cells. Press [Zero] to return to weighing. 6.2.4 F4 -INITIAL ZERO RANGE To enter this parameter, press the [Tare] key when “ ” is shown. The display will show the current initial zero range. © Adam Equipment Company 2007...
  • Page 16: F5 -Re-Zero Range

    ” is shown. The display will show if the successive tare is on or off. Press the [U.Wt./Units] key to change the value. Press [Tare] to accept the displayed value. Press [Zero] to return to weighing. © Adam Equipment Company 2007...
  • Page 17: F7 -A/D Count

    Press [Zero] to return to weighing. 6.2.9 F9 –PIN This parameter allows setting of password to enter these Service Parameters. To enter this parameter, press the [Tare] key when “ ” “ ” is shown. © Adam Equipment Company 2007...
  • Page 18: F10 -Low Voltage Detection

    ” is shown. The display will show if the LVD Mode is set to Press the [U.Wt./Units] key to change the value. Press [Tare] to accept the displayed value. Press [Zero] to return to weighing. © Adam Equipment Company 2007...
  • Page 19: Wiring Diagram

    7.0 WIRING DIAGRAM © Adam Equipment Company 2007...
  • Page 20: Mechanical Assembly

    8.0 MECHANICAL ASSEMBLY © Adam Equipment Company 2007...
  • Page 21: List Of Mechanical Assembly

    PCB Assy the load cell lower supporter the load cell the load cell upper supporter washer screw front display PCB Assy upper cover rear display overlay ABS pan SST pan © Adam Equipment Company 2007...
  • Page 22: Display Assembly

    9.0 DISPLAY ASSEMBLY © Adam Equipment Company 2007...
  • Page 23 Shielded interconnect cables must be employed with this equipment to insure compliance with the pertinent RF emission limits governing this device. Changes or modifications not expressly approved by Adam Equipment could void the user's authority to operate the equipment. WEEE COMPLIANCE...
  • Page 24 ADAM EQUIPMENT is an ISO 9001:2000 certified global organisation with more than 30 years experience in the production and sale of electronic weighing equipment. Products are sold through a world wide distributor network supported from our company locations in the UK, USA, SOUTH AFRICA and AUSTRALIA.