1) Switch on the scale then press the [kg/lb] key during the self-checking test of
the display.
2) The display will show the first function
3) Press the [kg/lb] key to select
first parameter.
4) Press the [Print] key to change the setting and [Zero] to save and move to the
next parameter, press [cm/ft] to exit.
To return to normal operation turn the scale OFF and then ON again.
6.9
TECHNICAL PARAMETERS
This is a menu available only to qualified service personnel for factory adjustments.
7.0

USER CALIBRATION

Switch on the scale and whilst the display counts down press the [Clear] key.
1)
2) The display will show kg or lb, press [Print] to select the unit of weight being
used for the calibration and then [Zero] to confirm.
3) The display will show CAL XX (XX being the last calibration weight to be used).
4) Press the [Print] key to change the value if required and [Zero] to confirm.
5) The display will show the Zero A-D which should be stable.
6) Press the [Zero] key to confirm the Zero point and the display will show LOd
XX (XX being the calibration weight value to be loaded onto the top pan).
7) Load the top pan with the calibration weight and press the [Zero] key.
8) The display will show "PASS" for a successful calibration and return to normal
weighing displaying the calibration weight value correctly.
9) Remove the weight from the pan and the scale will return to Zero.
*A print out with date and time can be issued following every calibration.
If an error code was displayed when calibrating please see the error code
table in section 9.1