Agilent Technologies Intuvo 900 Safety Manual

Agilent Technologies Intuvo 900 Safety Manual

Gas chromatograph

Advertisement

Quick Links

Agilent Intuvo 9000
Gas Chromatograph
Safety Manual
Agilent Technologies

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies Intuvo 900

  • Page 1 Agilent Intuvo 9000 Gas Chromatograph Safety Manual Agilent Technologies...
  • Page 2 Notices Warranty Safety Notices © Agilent Technologies, Inc. 2017 No part of this manual may be reproduced C A U T I O N The material contained in this in any form or by any means (including document is provided “as is,” and...
  • Page 3: Table Of Contents

    Agilent Intuvo 9000 Gas Chromatograph Safety Manual Introduction Important Safety Warnings Hydrogen Safety Uncombusted Hazardous Gases Electron Capture Detector (ECD) Safety and Regulatory Certifications Intended Use Cleaning Fume Hood Recycling the Product Agilent Technologies...
  • Page 4: Important Safety Warnings

    Introduction Important Safety Warnings Before moving on, there are several important safety notices that you should always keep in mind when using the Agilent GC. When handling/using chemicals for preparation or use within the GC, all WA RNING applicable local and national laboratory safety practices must be followed. This would include, but is not limited to, correct use of Personal Protective Equipment (PPE), correct use of storage vials, and correct handling of chemicals, as defined in the laboratory’s internal safety analysis and standard operating procedures.
  • Page 5 Introduction Electrostatic discharge is a threat to GC electronics The printed circuit (PC) boards in the GC can be damaged by electrostatic discharge. Do not touch any of the boards unless it is absolutely necessary. If you must handle them, wear a grounded wrist strap and take other antistatic precautions.
  • Page 6: Hydrogen Safety

    Introduction Hydrogen Safety Hydrogen gas may be used as carrier gas, and/or as fuel for the FID and NPD. When mixed with air, hydrogen can form fire hazard. Hydrogen is flammable. Leaks may create a fire hazard. In any application using WA RNING hydrogen, leak test all connections, lines, and valves before operating the instrument.
  • Page 7 Introduction • The small heated zones are turned off. • An alarm tone sounds. To recover from this state, fix the cause of the shutdown (tank valve closed, serious leak, others). Turn the instrument off, then back on. The GC cannot always detect leaks in inlet and/or detector gas streams. For this WA RNING reason, it is vital that column fittings should always be either connected to a column, or have a cap or plug installed.
  • Page 8: Uncombusted Hazardous Gases

    Introduction Uncombusted Hazardous Gases During normal operation of the GC with many detectors and inlets, some of the carrier gas and sample vents outside the instrument through the split vent, septum purge vent, and detector exhaust. If any sample components are toxic or noxious, or if hydrogen is used as the carrier gas, these exhausts must be vented to a fume hood.
  • Page 9: Electron Capture Detector (Ecd)

    Wear disposable plastic gloves and observe normal laboratory safety precautions. 2 Contact your local Agilent Technologies sales office or distributor for ECD disposal instructions 3 Include a letter stating the condition of abuse.
  • Page 10 For other countries, consult with the appropriate agency for equivalent requirements. Agilent Technologies recommends a vent line internal diameter of 6 mm (1/4-inch) or greater. With a line of this diameter, the length is not critical. Intuvo 9000 GC Safety Manual...
  • Page 11: Safety And Regulatory Certifications

    Introduction Safety and Regulatory Certifications The Agilent GC conforms to the following safety standards: • Canadian Standards Association (CSA): C22.2 No. 61010-1 • CSA/Nationally Recognized Test Laboratory (NRTL): ANSI/UL 61010-1 • International Electrotechnical Commission (IEC): 61010–1, 60101-2-010, 60101-2-081 • EuroNorm (EN): 61010–1 The Agilent GC conforms to the following regulations on Electromagnetic Compatibility (EMC) and Radio Frequency Interference (RFI): •...
  • Page 12 Your product must only use the power cord that was shipped with this product. Do not use this power cord with any other product. Information The Agilent Technologies Gas Chromatograph meets the following IEC (International Electro-technical Commission) classifications: Safety Class I, Transient Overvoltage Category II, Pollution Degree 2.
  • Page 13 Introduction protection of the Agilent Gas Chromatograph has been compromised, disconnect the unit from all power sources and secure the unit against unintended operation. Refer servicing to qualified service personnel. Substituting parts or performing any unauthorized modification to the instrument may result in a safety hazard.
  • Page 14 Failure to comply with these precautions violates safety standards of design and the intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
  • Page 15 4 Make sure that all peripheral devices are also certified. 5 Make sure that appropriate cables are used to connect the device to peripheral equipment. 6 Consult your equipment dealer, Agilent Technologies, or an experienced technician for assistance. 7 Changes or modifications not expressly approved by Agilent Technologies could void the user’s authority to operate the equipment.
  • Page 16: Intended Use

    Agilent Technologies Intended Use Agilent products must only be used in the manner described in the Agilent product user guides. Any other use may result in damage to the product or personal injury. Agilent is not responsible for any damages caused, in whole...