Inspection

Regular inspections must be carried out according
to the following specifications and at the stated
intervals. Technical documentation is available on
request.
WARNING
Risk of electric shock
Disconnect all electrical connections from the
power supplies before carrying out any work.
Component
Display mount and
arm system
Display mount and
arm system
38
Interval
Task
Before each
Functional and visual inspec-
use
tion, see page 18.
Every 12
Check accompanying docu-
months
ments: instructions for use
available.
Check equipment of the medi-
cal device for completeness
according to the instructions
for use.
Perform functional and visual
inspection.
Check for deformations of the
arm system and whether parts
have become loose.
Check for paint damage and
cracks in plastic parts.
Check whether all plastic parts
are present and in the correct
position.
Safety
Personnel
checks
responsible
X
User
X
Service per-
sonnel
X
X
X
Instructions for use Polaris Multimedia