Maintenance
Component
Display mount and
arm system
Safety checks
.
Checks
1)
Safety checks
1) Term applicable in the Federal Republic of Germany, corresponds to the "Periodic safety inspection" in Austria
Safety checks are not a substitute for maintenance,
which includes the preventive replacement of
wearing parts as specified by the manufacturer.
CAUTION
Risk of medical device failure
If the safety checks are not carried out regularly,
the functionality of the medical device may be
jeopardized.
Perform safety checks at the indicated intervals.
1 Check accompanying documents:
– Instructions for use available
2 Check that the product is in perfect condition:
– Labels complete and legible
– No deformation of the arm system and
loosening of parts
– No paint damage, cracks in plastic parts
– Check the arm system (central axis, swivel
arm, spring arm) for collision damage
(welding seams free of cracks)
3 Check the components and accessories for
completeness according to the instructions for
use.
40
Interval
Task
Every 10
Electrical safety inspection of
years
the display mounting system
according to IEC 62353.
Replace components if neces-
sary.
Interval
see tables "Inspec-
tion" and "Preven-
tive maintenance"
Safety
checks
Personnel responsible
see tables "Inspection" and "Preventive
maintenance"
4 Check electrical safety according to IEC 62353.
5 Check safety features:
– Dimensions of the locking element (at the
spring arm). The display mount and the arm
system must be de-energized.
– Check the position and form of the circlip of
the arm system. The display mount and the
arm system must be de-energized.
Instructions for use Polaris Multimedia
Personnel
responsible
X
Specialized
service per-
sonnel