Operators Maintenance

WA RN ING
CAU TI O N
Installation and Verification Manual
No operator serviceable parts inside. Refer servicing to qualified
personnel. To prevent electrical shock, do not remove covers.
Maintenance appropriate for the operator is:
• Cabinet cleaning - clean the cabinet using a dry cloth only.

• Cleaning Optical Connectors

• Ensure ventilating fan cover is clean.
Cleaning Optical Connectors
You should clean the optical connectors at regular intervals
using the following materials:
Description
Compressed Air Can or Blow Brush
Isopropyl Alcohol
Lens Cleaning Paper
Swabs
Do not insert any tool or object into the optical IN or OUT ports of the
instrument as damage to or contamination of the optical fiber may
result.
To clean the optical connectors
1 Disconnect the instrument from the Power Line or switch off
the laser transmitter before commencing this cleaning
procedure.
2 Remove the adapters from the optical IN and OUT ports by
flipping back the lever on the optical adapter.