Input A, B, C, D DC Voltage Accuracy Test

To prevent possible electrical shock, fire, or personal injury,
ensure that the calibrator is in standby mode before making any
connection between the calibrator and the Test Tool.
Dangerous voltages are present on the calibration source and
connection cables during these steps.
The test steps for channel C and D are only for the models
190-104, 190-204, and 190-504.
Proceed as follows to verify the automatic dc voltage scope measurement:
1. Connect the Test Tool to the 5502A as shown in see Figure 4-4.
PM9092
PM9093
A
METER
B
A
2 CHANN. + METER
4 CHANNELS
Figure 4-4. Test Tool Inputs A, B, C, D to 5502A Normal Output
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
 Warning
Note
PM9091
PM9092
CONNECT TO CHANNEL A, B, C, D IN PARALLEL
NO
50 OHM TERMINATION
B
C
D
NORMAL
PM9081
FLUKE 5502A CALIBRATOR
Performance Verification
Scope Input A, B, C, D Tests
SCOPE
4
perf-ver-b2.eps
4-11