Fluke ScopeMeter 190 II Series Service Manual page 122

Hide thumbs Also See for ScopeMeter 190 II Series:
Table of Contents

Advertisement

To prevent possible electrical shock, fire, or personal injury,
ensure that the calibrator is in standby mode before making any
connection between the calibrator and the Test Tool.
Dangerous voltages are present on the calibration source and
connection cables during these steps.
4. Set the 5502A to operate (OPR).
F3
5. Press
6. Wait until the display shows calibration status
F2
7. Press
calibration point, and start the calibration. Continue through all calibration
points of Table 5-17.
8. Set the 5502A to Standby, and continue at the Input ABCD (AB) Position
section.
Table 5-17. Input ABCD Gain Calibration Points
Cal step
CL 0660
CL 0604
CL 0637
CL 0504
CL 0624
CL 0599
CL 0600
CL 0601
CL 0602
CL 0622
CL 0662
CL 0605
CL 0606
CL 0607
CL 0664
CL 0608
CL 0609
CL 0610
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
 Warning
to start the calibration.
to select the next calibration step, set the 5502A to the next
UUT input value (5502A SCOPE, MODE volt, SCOPE Z 1 MΩ, 1 kHz)
300 mV
500 mV
none
(5500 standby)
500 mV
none
(5500 standby)
10 mV
25 mV
50 mV
100 mV
none
(5500 standby)
2 V
1 V
2.5 V
5 V
20 V
10 V
25 V
50 V
(set 5502A to OPR)
Calibration Adjustment
Final Calibration (Firmware: V10.9 and Lower)
.
:READY
5
5-37