Fluke ScopeMeter 190 II Series Service Manual page 34

Hide thumbs Also See for ScopeMeter 190 II Series:
Table of Contents

Advertisement

 Power
Fluke 190-xx4, -50x: Rechargeable Li-ion Battery (model BP291):
Operating Time ................................................... up to 7 hours (Low Intensity)
Charging Time ..................................................... 5 hours
Capacity/Voltage ................................................. 52 Wh / 10.8 V
Fluke 190-062, -102, -202: Rechargeable Li-ion Battery (model BP290):
Operating Time .................................................... up to 4 hours (Low Intensity)
Charging Time ..................................................... 2.5 hours
Capacity/Voltage ................................................. 26 Wh / 10.8 V
Rechargeable Li-ion Battery (model BP 290 and BP291):
Life Time (>80 % capacity) .................................. 300 x charge/discharge
Allowable ambient temperature
during charging ....................................................... 0 °C to 40 °C (32 °F to 104 °F)
Auto power down time
(battery saving) ................................................... 5 min, 30 min or disabled
Power Adapter BC190:
BC190/801 European line plug 230 V ±10 %
BC190/813 North American line plug 120 V ±10 %
BC190/804 United Kingdom line plug 230 V ±10 %
BC190/806 Japanese line plug 100 V ±10 %
BC190/807 Australian line plug 230 V ±10 %
BC190/808 Universal switchable adapter 115 V ±10 % or 230 V ±10 %, with plug EN60320-2.2G
BC190/820 Universal adapter 100...240 V ±10 %, with plug EN60320-2.2G
Line Frequency ........................................................ 50 Hz and 60 Hz
Probe Calibration
Manual pulse adjustment and automatic DC adjustment with probe check
Generator Output .................................................... 1.225 Vpp / 500 Hz square wave
Internal Memory
Number of Scope Memories .................................... up to 30 (varies according to version)
Each memory can contain 2/4 waveforms plus corresponding setups
Number of Recorder Memories ............................... 10
Each memory can contain:
• 2/4 channel input TrendPlot
• 2/4 channel input Scope Record
• 100 2/4 channel input Scope screens (Replay)
Number of Screen Image memories ....................... 9
Each memory can contain one screen image
External Memory
USB stick, 2GB max
Mechanical
Size ......................................................................... 265 mm x 190 mm x 70 mm (10.5 in x 7.5 in x 2.8 in)
Weight
Fluke 190-xx4 ...................................................... 2.2 kg (4.8 lb) including battery
Fluke 190-5xx ...................................................... 2.2 kg (4.8 lb) including battery
Fluke 190-xx2 ...................................................... 2.1 kg (4.6 lb) including battery
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2
Specifications
Miscellaneous
2-13

Advertisement

Table of Contents
loading

Table of Contents