Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
Interface Ports
Two USB ports provided. Ports are fully insulated from instrument's floating measurement circuitry:
•
A USB-host port directly connects to external flash memory drive ('USB-stick', ≤2 GB) for storage of waveform data,
measurement results, instrument settings and screen copies.
•
A mini-USB-B is provided which allows for interconnection to PC for remote control and data transfer using SW90W
(FlukeView
software for Windows
•
One port can be active at the same time, so remote control and data transfer via mini-USB is not possible when
saving or recalling data to or from the USB-stick.
Environmental
Environmental ......................................................... MIL-PRF-28800F, Class 2
Temperature
Operating ............................................................ 0 °C to 50 °C (32 °F to 122 °F)
Operating and charging ...................................... 0 °C to 40 °C (32 °F to 104 °F)
Storage ............................................................... -20 °C to +60 °C (-4 °F to +140 °F)
Humidity (Maximum Relative)
Operating
0 °C to 10 °C (32 °F to 50 °F) ......................... noncondensing
10 °C to 30 °C (50 °F to 86 °F) ....................... 95 % (± 5 %)
30 °C to 40 °C (86 °F to 104 °F) ..................... 75 % (± 5 %)
40 °C to 50 °C (104 °F to 122 °F) ................... 45 % (± 5 %)
Storage
-20 °C to +60 °C (-4 °F to +140 °F) ................ noncondensing
Altitude
Operating
CATIII 600 V, CATII 1000 V ............................ 3 km (10,000 feet)
CATIV 600 V, CATIII 1000 V .......................... 2 km (6,600 feet)
Storage ............................................................... 12 km (40,000 feet)
Vibration (Sinusoidal) .............................................. max. 3 g
Vibration (Random) ................................................. 0.03 g
Shock ...................................................................... max. 30 g
Electromagnetic Environment ................................. EN/IEC61326-1(Portable Equipment)
Enclosure Protection ............................................... IP51, ref: IEC60529
Certifications
Conforms to ............................................................
Safety
Designed for 1000 V Measurement Category III, 600 V Measurement Category IV (with supplied 10:1 probes) in
accordance with:
•
EN/IEC 61010-1, Pollution Degree 2
•
EN/IEC 61010-2-030
•
IEC 61010-031
BNC Input A, B, (C, D) directly ............................... 300 V CAT IV
Via VPS410 ............................................................. 1000 V CAT III, 600 V CAT IV
METER/EXT banana input ..................................... 1000 V CAT III, 600 V CAT IV
2-14
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
).
2
/Hz
Max. Input Voltages