10:1 Probe Vps410; Electromagnetic Immunity - Fluke ScopeMeter 190 II Series Service Manual

Hide thumbs Also See for ScopeMeter 190 II Series:
Table of Contents

Advertisement

Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual

10:1 Probe VPS410

Accuracy
Probe accuracy when adjusted on the test tool:
DC to 20 kHz....................................................... ±1 %
20 kHz to 1 MHz ................................................. ±2 %
1 MHz to 25 MHz ................................................ ±3 %
For higher frequencies the probe's roll-off starts affecting the accuracy.
For further probe specifications see the instruction sheet supplied with the VPS410 probe set.

Electromagnetic Immunity

The Fluke 190 Series II test tools, including standard accessories are compliant to EN 61326-1, with the addition of the
table that follows.
Scope Mode (10 ms/div: Waveform disturbance with VPS410 voltage probe shorted (see table below).
Frequency
80 MHz – 450 MHz
450 MHz – 1 GHz
1.4 GHz – 2 GHz
2 GHz – 2.7 GHz (1 V/m)
2-16
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
(E = 3V/m)
Disturbance <10 % of
No Disturbance
full scale
≥500 mV/d
100, 200 mV/div
All ranges
All ranges
All Ranges
Disturbance >10 % of
full scale
2, 5, 10, 20, 50 mV/div