Fluke ScopeMeter 190 II Series Service Manual page 102

Hide thumbs Also See for ScopeMeter 190 II Series:
Table of Contents

Advertisement

Table 5-3. Input C LF-HF Gain Calibration Points
Cal step
UUT input signal
CL 0694
0.5 Vpp square wave, 1 kHz
CL 0455
100 mVpp square wave, 1 kHz
CL 0550
2.5 Vpp square wave, 1 kHz
CL 0584
2.5 Vpp sine wave, 50 kHz
CL 0585
2.5 Vpp sine wave
Red subversion 5; Blue Subversion 2
Fluke 190-502/504:
Fluke 190-202/204:
Fluke 190-102/104:
Fluke 190-062:
CL 0484
500 mVpp sine wave, 50 kHz
CL 0485
0.5 Vpp sine wave
Red subversion 5; Blue Subversion 2
Fluke 190-502/504:
Fluke 190-202/204:
Fluke 190-102/104:
Fluke 190-062:
Fluke 190-202/204:
Fluke 190-102/104:
Fluke 190-062:
CL 0464
100 mVpp sine wave, 50 kHz
CL 0465
100 mVpp sine wave
Red subversion 5; Blue Subversion 2
Fluke 190-502/504:
Fluke 190-202/204:
Fluke 190-102/104:
Fluke 190-062:
Fluke 190-202/204:
Fluke 190-102/104:
Fluke 190-062:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SCOPE edge, 0.5 Vpp, 1 kHz
SCOPE edge, 100 mVpp, 1 kHz
SCOPE edge, 2.5 Vpp, 1 kHz
SCOPE levsine, 2.5 Vpp, 50 kHz
SCOPE levsine, 2.5 Vpp,
501 MHz
501 MHz
381 MHz
381 MHz
151 MHz
151 MHz
111 MHz
111 MHz
SCOPE levsine, 500 mVpp, 50 kHz
SCOPE levsine, 0.5 Vpp,
501 MHz
501 MHz
381 MHz
381 MHz
151 MHz
151 MHz
111 MHz
111 MHz
221 MHz
221 MHz
141 MHz
141 MHz
91 MHz
91 MHz
SCOPE levsine, 100 mVpp, 50 kHz
SCOPE levsine, 100 mVpp,
501 MHz
501 MHz
381 MHz
381 MHz
151 MHz
151 MHz
111 MHz
111 MHz
221 MHz
221 MHz
141 MHz
141 MHz
91 MHz
91 MHz
Calibration Adjustment
Final Calibration For v11.10 and later
5502A Setting
5
5-17