Equipment Requirements For Verification; General Instructions; Operation Instructions; Reset The Test Tool - Fluke ScopeMeter 190 II Series Service Manual

Hide thumbs Also See for ScopeMeter 190 II Series:
Table of Contents

Advertisement

Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual

Equipment Requirements for Verification

The primary source instrument used in the verification procedures is the Fluke
5502A. If a 5502A is not available, you can substitute another calibrator as long
as it meets the minimum test requirements.
Fluke 5502A Multi-Product Calibrator, including SC600 Oscilloscope
Calibration Option.
Stackable test leads (4x) as supplied with the 5502A.
50 Ω Coax Cables (2x): use Fluke PM9091 (1.5 m, 3 pcs./set) and PM9092
(0.5 m, 3 pcs./set).
Male BNC to Dual Female BNC adapter (1x), Fluke PM9093/001.
50 Ω feed through termination, always use Fluke TRM50 for Fluke 190-502
and 190-504.
Dual Banana Plug to Female BNC Adapter (1x), Fluke PM9081/001.
Dual Banana Jack to Male BNC Adapter (1x), Fluke PM9082/001.
TV Signal Generator (part of SC600 600 MHz Oscilloscope Calibration
Option).
10:1 Attenuator Probes as supplied with Test Tool.

General Instructions

Follow these general instructions for all tests:
Power the Test Tool with the BC190 power adapter. The battery pack must
be installed.
Allow the 5502A to satisfy its specified warm-up period.
For each test point, wait for the 5502A to settle.
Allow the Test Tool a minimum of 30 minutes to warm up.
One division on the LCD consists of 25 pixels (1 pixel = 0.04 division).
This procedure is set up for all models of the Test Tool. These have either
two oscilloscope channels A and B with BNC inputs and a multimeter channel
with banana inputs, or four oscilloscope channels.
The figures that show the connection between calibrator and Test Tool are
universal and also show the connection between calibrator and a Test Tool
with four oscilloscope channels (for instance model 190-204).

Operation Instructions

Reset the Test Tool

Proceed as follows to reset the Test Tool:
1. Press
2. Press and hold
3. Press and release
4. Wait until the Test Tool beeps twice and then release
indicate a successful reset.
4-4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
to turn off the Test Tool.
.
USER
to turn on the Test Tool.
. Two beeps
USER