Input ABCD (AB) Position

To do the Input ABCD (AB) Position calibration:
1. Press
2. Remove all Input A, B, C, D (A, B) connections (open inputs).
3. Press
4. Wait until the display shows calibration status
5. Continue at the Input ABCD (AB) Volt Gain section.

Input ABCD (AB) Volt Gain

To prevent possible electrical shock, fire, or personal injury,
ensure that the calibrator is in standby mode before making any
connection between the calibrator and the Test Tool.
Dangerous voltages are present on the calibration source and
connection cables during these steps.
The adjustment steps for channel C and D are only for the models
190-104 and 190-204.
To do the Input ABCD (AB) Volt Gain calibration:
1. Press
2. Connect the Test Tool to the 5502A as shown in Figure 5-16.
PM9093
PM9092
METER
A
2 CHANN. + METER
3. Set the 5502A to supply a DC voltage (NORMAL output) to the first
calibration point in Table 5-18.
4. Set the 5502A to operate (OPR).
5. Press
6. Wait until the display shows calibration status
5-38
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
F2
to select calibration adjustment step
F3
to start the calibration.
 Warning
F2
to select the first calibration step in Table 5-18.
PM9091
PM9092
B
A
B
C
D
4 CHANNELS
PM9081
Figure 5-16. Test Tool Input ABCD to 5502A Normal Output
F3
to start the calibration.
CL 0619.
:READY.
Note
CONNECT TO CHANNEL A, B, C, D IN PARALLEL
NO
50 OHM TERMINATION
NORMAL
SCOPE
FLUKE 5502A CALIBRATOR
.
:READY

Advertisement

Table of Contents
loading

Table of Contents