Save Calibration Data and Exit

To save the calibration data and exit the Maintenance mode:
1. Remove all test leads from the Test Tool inputs.
F4
2. Press
Calibration data valid.
Save data and exit maintenance mode?
Calibration data valid indicates that the calibration adjustment
procedure is performed correctly. It does not necessarily mean that
the Test Tool meets the specifications listed in Chapter 2.
F4
3. Press
After saving the calibration data, the calibration number and date updates if
the calibration data changes and the data are valid.
The calibration number and date do not change if:
the calibration mode is entered and left without doing a calibration
adjustment.
only the contrast calibration adjustment and/or the probe calibration are
done.
F3
If you press
You can either calibrate the Test Tool again, or press
F4
(
to save and exit.
YES)
Error messages:
WARNING: Calibration data not valid.
Save data and exit maintenance mode?
If you did the Warming-Up and Pre-Calibration successfully and you want to
store the Pre-Calibration data before continuing with the Final Calibration:
F4
1. Press
When you turn off and turn on the Test Tool again, it will show the message:
The instrument needs calibration.
Please contact your service center.
The calibration date and number do not update. You must continue with the
Final Calibration.
To return to the Maintenance mode and repeat the complete calibration:
F3
1. Press
F1
2. Press
calibrate the Test Tool, starting at Warming-Up and Pre-Calibration.
If you want to exit and maintain the old calibration data:
1. Turn the Test Tool off.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
(
. The Test Tool shows on the display:
EXIT)
Note
(
to save and exit.
YES)
Note
(
, the Test Tool returns to the calibration mode.
NO)
(YES).
(NO).
until the display shows
WarmingUp (CL 0200):IDLE
Calibration Adjustment

Save Calibration Data and Exit

F4
(
,
EXIT)
, and
5
5-43