Meter Tests
The following tests are for the models 190-062, 190-102,
190-202, and 190-502.
Meter DC Voltage Accuracy Test
To prevent possible electrical shock, fire, or personal injury,
ensure that the calibrator is in standby mode before making any
connection between the calibrator and the Test Tool.
Dangerous voltages are present on the calibration source and
connection cables during these steps.
Proceed as follows to test the meter dc voltage measurement accuracy:
1. Connect the Test Tool to the 5502A as shown in Figure 4-13.
2. Select the Test Tool setup.
METER
3. Press
F1
4. Press
5. Press
ENTER
.
MANUAL
6. Press
AUTO
to select MANUAL ranging. Use
ranges.
7. Set the range to the first test point in Table 4-17.
8. Set the 5502A to source the appropriate dc voltage.
9. Observe the reading and check to see if it is within the range shown under
the appropriate column.
10. Continue through the test points.
11. When you are finished, set the 5502A to 0 (zero) Volt and Standby.