Fluke ScopeMeter 190 II Series Service Manual page 110

Hide thumbs Also See for ScopeMeter 190 II Series:
Table of Contents

Advertisement

A
METER
B
2 CHANN. + METER
Figure 5-9. 5502A NORMAL Output to Test Tool Banana Input
3. Set the 5502A to supply a DC voltage to the first calibration point in
Table 5-7.
4. Set the 5502A to operate (OPR).
F3
5. Press
to start the calibration.
6. Wait until the display shows calibration status
F2
7. Press
to select the next calibration step, set the 5502A to the next
calibration point, and start the calibration. Continue through all calibration
points of Table 5-7.
8. Set the 5502A to Standby, and continue at the Multimeter (DMM) Ohm Gain
section.
Table 5-7. DMM Volt Gain Calibration Points
Cal step
CL 0840
CL 0849
CL 0841
CL 0842
CL 0843
CL 0844