150X Series
Calibration Manual
Step
Voltage Range
1
2
3
4
5
6
7
8
9
Insulation Function, External Sense
The following test verifies that the Tester will sense a voltage > 30 V when present on the
circuit under test.
Source Voltage Accuracy Test, "R" Nominal
Complete the test steps in Table 6 to verify source voltage of the insulation function
under load. If using the Fluke 5320A, use the high resistance function to verify source
voltage.
18
Table 5. Insulation Resistance Accuracy Test
Insulation
Applied
1000 V
9 GΩ
1000 V
1.9 GΩ
1000 V
1 MΩ
1000 V
49 MΩ
1000 V
60 MΩ
500 V
500 kΩ
250 V
250 kΩ
100 V
100 kΩ
50 V
50 kΩ
1. Connect the UUT
INSULATION
voltage output terminals.
2. Set the 5320A to the voltage output function (V).
3. Turn the UUT rotary function switch to an insulation function.
4. Apply 35 V, 50 Hz to the UUT.
5. Verify that the UUT displays > 30 V in the primary display, and the red LED
lightning bolt comes on.
1. Connect the Fluke 5320A high resistance output terminals to the UUT
and
INSULATION
COM
2. Put the Fluke 5320A in the high resistance function (H).
3. Set the Fluke 5320A for the applied load shown in Table 6 for Steps 1-5.
4. Press T and verify that the UUT and Fluke 5320A voltage readings are
within the limits of Table 6. Record both of these readings.
5. Using the Fluke 5320A voltage reading as the reference, calculate the UUT
voltage reading error %, (5320A V - UUT V / 5320A V) x 100, and record for
later use.
riešenia na presné meranie
Display Units
MΩ
and
output terminals to the Fluke 5320A
COM
terminals.
http://www.elso.sk
Display Reading Limits
1503
1507/1508
N/A
7.8 to 10.2
1862 to 1938
N/A
0.5 to 1.5
0.5 to 1.5
47.5 to 50.5
47.8 to 50.2
58.3 to 61.7
58.6 to 61.4
0.44 to 0.56
0.44 to 0.56
N/A
0.20 to 0.30
N/A
0.6 to 0.15
N/A
0.0 to 0.10