66 . Lambda 365 Users Guide
Tungsten Lam p M odule Replacem ent
1.
Make sure that the main power is turned off. If the system and lamp(s) were turned
on at least 20 minutes should pass before the lamp changing process is started. This
time allows the lamps to cool and return to room temperature.
Use the Tungsten Lamp Changing Tool (Part No. N4102048) when
replacing the Tungsten Lamp. In addition a pair of appropriate
disposable gloves should be worn during the lamp replacement
process. Please avoid touching the glass bulb section of the lamp this
CAUTION
can leave oil deposits on the bulb which can drastically reduce lamp
life.
Utilisez Tungsten Lamp outil Modification lors du remplacement de la
ATTENTION
lampe de tungstène. En outre, une paire de gants jetables appropriés
doivent être portés pendant le processus de remplacement de la
lampe. S'il vous plaît éviter de toucher la section de l'ampoule de verre
de la lampe, cela peut laisser des dépôts d'huile sur le bulbe qui peut
réduire considérablement la vie de la lampe.
Tungsten Lamp Changing Tool (Part No. N4102048)
2.
Remove the white bolt covers (four each) using screwdriver and remove the M3*12
screws (four each) using a Phillips screwdriver to open the vent cover.