In the summary in the figure all segments
•
segment number,
•
temperature, CO
Confirms and saves the changes
Cancels the entered changes to segments and goes into program parameters.
With more parameters you can scroll the panel up and down.
The option is available if the number of segments is equal to 2 or more.
When the program finishes the last segment, the unit starts the program again from the first segment.
You can define if a program should be carried out once (loop: 1) or multiple times (loop: 2 to 255). You can also set
the program to be carried out continuously, tick the „∞" option.
Can be set in terms of:
Parameters:
The program without a ramp – the unit starts the countdown of the segment time when the set temperature is reached.
The program with a ramp – first, the unit counts down the time of the ramp and then proceeds to the segment countdown when
the set temperature is reached. Regardless of whether the time of ramp elapsed.
27
vwr.com I Instruction manual VWR
Summary of segments
(Figure 16)
concentration, duration time, target time of reaching temperature of a given segment.
2
Figure 16 - The summary of the program
Loop
Priority
®
CO
incubator
2
are visible along with the entered settings: