This modification can be performed during operation. We do recommend, however, to
switch off the device during the modification.
4
Operation
During operation, take care that the device remains within its intended pressure and tem-
perature ranges. No other monitoring is necessary.
Permissible ambient temperature:
4.1
Test Terminals
You can check the output current without interrupting the current loop, using the test ter-
minals on the terminal board. When you connect a current meter to the "+Test" and "-Test"
terminals, the current is automatically routed through it.
Figure 4: Current measurement during operation using test terminals
BA_078 Rev 1F6 Level transmitter PASCAL Ci4 LEVEL
Figure 3: Remote display and control unit after installation
A
-40...80 °C
Page 5/36