Page 1
4.2 Devices with Diaphragm Seal ................. 4 4.3 Maintenance / Service ..................... 4 5 Disassembly ........................5 LABOM Mess- und Regeltechnik GmbH Im Gewerbepark 13 27798 Hude Germany BA_032_en_2021-11_10.01 Hotline: +49 4408 804-444 Fax: +49 4408 804-100 e-mail: sales@labom.com www.labom.com...
Page 2
In addition to this instruction, be sure to observe all statutory requirements, applica- ble standards, the additional technical specifications on the accompanying data sheet (see www.labom.com) as well as the specifications indicated on the type plate. General Safety Notes The installation, set up, service or disassembly of this device must only be done by trained, qualified personnel using suitable equipment and authorized to do so.
Page 3
(see figure 1). You can find further information about zero- adjustment of pressure gauge with micro adjustment pointer in the document TA_029 on www.labom.com. Figure 1: Zero-point correction BA_032_en_2021-11_10.01 Pressure transmitter with bourdon tube Page 3/5...
Page 4
4.1.2 Adjusting the angle-of-rotation sensor For adjustable devices the adjustment re- quires to apply the lower limit pressure re- spectively the upper limit pressure to the de- vice. Bring the enclosed or another magnet in the alignment position (see picture) to ad- just accordingly.
Page 5
Disassembly When measuring hot media, make sure that the device has cooled down prior to any dis- mounting or wear appropriate protective clothing to avoid burns. Switch off the power supply to the device before disconnecting the electrical connections. Once this is done, the device may be mechanically removed. Warning Opening pressurized lines might cause severe injuries.