Display Mode / Measured Value Display - Labom CI4 Series Operating Instructions Manual

Level transmitter pascal ci4 level
Hide thumbs Also See for CI4 Series:
Table of Contents

Advertisement

− Header: Icon description, if applicable. Otherwise "Value"
− Data area: Measured values and parameters according to the selected display
mode (see 6.3.4)
Display of device data (see 6.3.1)
− Header: Title for the displayed device data
− Data area: Device data
Operating menue
− Header: Selected menue item
− Data area: Submenue or dialogue during setting procedure
6.3

Display Mode / Measured Value Display

After applying the supply voltage and completion of the initialisation, the device switches to
the display mode. In the display, the currently measured value is shown.
If the measured value is larger than the displayable number due to setting a fixed decimal
point, the largest displayable number is shown flashing. In this case please select a differ-
ent setting of the decimal point.
In display mode you can review the device data, enter the menue or lock/unlock the
menue.
Button
Function
/
Scroll through the pages with device data
OK
Go to the operating menue
ESC
Return to measured value display
ESC+OK Activate / deactivate menue lock (see 6.3.2)
6.3.1
Quick Access to Device Data
You can access a number of device parameters directly from the measured value display
using the
/
-buttons. This allows a quick overview of the device configuration.
With the
-button you can display variable data (trailing pointers and counters), with the
-button static information such as configuration data.
From any screen of the device data, you can enter the operating menue with OK and go
back to the measured value display by pressing ESC.
The sequence of the screens with device data is as follows:
Counter (operating hour counter, maintenance timer)
Min/Max-values (fill height and sensor temperature)
− ----- Measured value display (starting point) ---
Pressure measurement (nominal range, damping, measuring rate)
Fill level measurement (measurand, density, height offset etc.)
Current output (characteristic curve, alarm current, limits)
BA_078 Rev 1F6 Level transmitter PASCAL Ci4 LEVEL
Table 2: Button functions in display mode
Page 12/36