Problem
Intermittent alarm
condition, and the (red)
alarm light illuminates
intermittently.
Intermittent alarm
condition, and the (red)
alarm light illuminates
continuously.
Unit does not start on
battery power.
Delay in recharging
internal battery.
EN-40
Probable Cause
Breathing rate has
exceeded the capacity
of the FreeStyle unit.
A general malfunction
has occurred.
Unit may be hot or cold
if left outdoors such as
in an automobile.
Internal battery
exceeds charging
temperature.
FreeStyle™
Solution
Reduce activity, and then
turn unit off and back on
again to reset unit. If
necessary, change to
another source of oxygen
as available and contact
your Equipment Provider.
Change to another
source of oxygen as
available, and contact
your Equipment Provider.
Allow the unit to reach
normal operating
temperature, which may
take several minutes if
exposed to temperature
extremes. Temporarily
connect your Universal
AC/DC power supply to
the unit's power inlet and
power source, as
needed, to reset the
unit's internal battery.
Unit may be operated;
however, charging may
not resume until battery
temperature is reduced.
(See page 9 notes.)