Summary of Contents for AirSep Newlife Intensity 10

  • Page 1 MANUEL D’UTILISATION GEBRAUCHSANWEISUNG MANUAL DEL USUARIO ΟΔΗΓΙΕΣ ΧΡΗΣΗΣ USER’S MANUAL MANUALE DI UTILIZZAZIONE HANDLEIDING MANUAL DE UTILIZAÇÃO HANDBOK BRUGERVEJLEDNING...
  • Page 2 NewLife Intensity 10 ® MN135-1 E...
  • Page 3: Table Of Contents

    NewLife Intensity 10 ® CONTENTS GENERAL SAFETY GUIDELINES ............... 1-GB I. DESCRIPTION ..................2-GB I. 1. Front panel (Fig. I. 1) ................2-GB I. 2. Rear panel (Fig. I. 2) ................2-GB II. STARTING-UP / INSTALLATION ............ 3-GB II. 1. Use in direct oxygen therapy ..............3-GB III.
  • Page 4 NewLife Intensity 10 ® Fig. I.1 Fig. I.2 MN135-1 E...
  • Page 5: General Safety Guidelines

    • Do not use in a specifically magnetic environment (MRI, etc.). This user’s manual reflects the instruction and safety guidelines for the “user” of the equipment, which AirSep acknowledges may be referred to as “patient,” “client,” or some other related term throughout various parts of the world.
  • Page 6: Description

    NewLife Intensity 10 ® - The device is used in accordance with the instructions for use.” If the replacement parts used for the periodic servicing by an approved technician do not comply with the manufacturer’s specifications, the latter is absolved from all responsibility in the event of an accident. Do not open the device whilst in operation: risk of electrical shock.
  • Page 7: Starting-Up / Installation

    NewLife Intensity 10 ® II. STARTING UP / INSTALLATION II. 1. Use in direct oxygen therapy. a - Ensure that the switch (1) is in the 0 position. b - If used with a humidifier: Unscrew the flask and fill it with water up to the line (see the humidifier instructions).
  • Page 8: Cleaning - Maintenance

    NewLife Intensity 10 ® III. CLEANING - MAINTENANCE III. 1. Cleaning Only the outside of the NewLife Intensity 10 is to be cleaned, with a dry cloth or, if ® necessary, a damp sponge and clean or soapy water, then thoroughly dried with wipes and an alcohol-based solution.
  • Page 9: Useful Information

    NewLife Intensity 10 ® IV. USEFUL INFORMATION IV. 1. Accessories and spare parts The accessories used with the NewLife Intensity 10 must: ® - be oxygen compatible, - be biocompatible, - comply with the general requirements of the 93/42/EEC European Directive. The connectors, tubes, nasal cannulas, probes or masks must be designed for oxygen therapy.
  • Page 10: 4. Alarms - Safety Devices

    NewLife Intensity 10 ® IV. 4. Alarms - Safety devices IV. 4. 1. Alarms • No volt detection: In the event of a mains power cut, a continuous audible alarm is tripped. • Operating fault: In the case of a distribution fault, an audible intermittent alarm is tripped. IV.
  • Page 11: 6. Technical Characteristics

    NewLife Intensity 10 ® IV. 6. Technical characteristics Dimensions: D x W x H: 368 x 419 x 699 mm. Weight: 26.4 kg. Noise level: ±55 dBA Flow values: 0 - 10 l/min. Average oxygen content: • at 2 l/min. 92%. •...
  • Page 12: 7. Standards

    NewLife Intensity 10 ® IV. 7. Standards EN 60-601-1-2: Electromagnetic compatibility of electro-medical devices. ISO 13485: Medical Devices Quality Management System. IV. 8. Symbols - Abbreviations : ON : Off (power switched off). : Type BF Applied Part (degree of protection against electric shock) : Class II device : Do not smoke.
  • Page 13: 11. Troubleshooting

    NewLife Intensity 10 ® IV. 11. Troubleshooting Observations Probable causes Solutions The 0-I button is in the ON Power cable not plugged in Check the cable connection. position. The device does not correctly. Mains power failure. Reset the circuit breaker (2) if operate.
  • Page 14: Appendices

    NewLife Intensity 10 ® APPENDICES INSTALLATION CHECK LIST / BEFORE USE Each time after installation and before use, carry out the following sequence of operations • Check that the device is in good condition generally (no traces of impacts, etc.) •...
  • Page 15 NewLife Intensity 10 ® European Representative: Medical Product Services GmbH Borngasse 20 35619 Braunfels, Germany info@mps-gmbh.eu MN135-1 E 11-GB...
  • Page 16: Dual Flow And Pediatric/Low Flow Options

    NewLife Intensity 10 ® Dual Flow and Pediatric/Low Flow Options The following information will explain the 10 liter dual flow and pediatric/low flow options for the NewLife ® Intensity 10 Oxygen Concentrator. Dual Flow Application The NewLife Intensity 10 unit’s 10-liter dual flow option allows a single concentrator to meet ®...
  • Page 17 NewLife Intensity 10 ® MN135-1 E...
  • Page 18 NewLife Intensity 10 ® MN135-1 E...
  • Page 19 CAIRE Inc 2200Airport Industrial Dr., Ste. 500 Ball Ground, GA 30107 USA www.chartindustries.com/RespiratoryHealthcare |MN135-1_Ea~ MN135-1_E...