AirSep FreeStyle Manual page 226

Hide thumbs Also See for FreeStyle:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 215
EN
When connected to AC or DC power, the unit's battery charges
until it reaches full capacity, either while the unit is operating or
turned off.
If the internal battery fully depletes and the FreeStyle unit shuts
down, the unit cannot be restarted with the AirBelt. Should this
occur, connect your FreeStyle to its Universal AC/DC power
supply for a short period of time in order to provide sufficient
internal battery power to start the unit. AirBelt can then be
connected to provide additional run time.
If the FreeStyle power supply remains connected when the
battery is fully charged, the four LEDs will turn off within
approximately 1 hour.
It may be necessary to initially connect the Universal AC/DC
FreeStyle power supply to the FreeStyle unit before the unit will
operate for the first time on battery power. Your Equipment
Provider may have already performed this step for you.
Use only AirSep Part No. FI144-1 for air intake filter for this unit.
Do not operate FreeStyle without the air intake filter in place. If a
second filter is provided, insert the "replacement" filter before
you clean the dirty filter.
AirSep does not recommend the sterilization of this
equipment.
Do not attempt any maintenance other than the possible
solutions listed within this manual.
EN-10
FreeStyle™
MN131-1 F

Advertisement

Table of Contents
loading

Related Products for AirSep FreeStyle

Table of Contents