System Power Source Power Requirements:
ST 115 ......... Operates on 97-127 VAC, 50/60Hz
120 Watts maximum at 115 VAC, 60Hz
ST 115E ......Operates on 197-253 VAC 50/60Hz
120 Watts maximum at 230 VAC, 50Hz
Temperature Specifications:
Tip Temperature Range:
Temperature Stability:
Actual minimum and maximum Operating Tip Temperatures may vary depending on
Handpiece, Tip selection and application.
Vacuum And Air
Measurements at front panel Vacuum and Controllable Pressure Ports.
Vacuum Rise Time: ------- 200 ms average as measured by PACE
Vacuum: ............................ 51 cm Hg. (20 in. Hg.) (nominal)
Pressure: .......................... 1.44 Bar (21 P.S.I.) (nominal at MAX setting)
Air Flow: ........................... 6 SLPM (0.22 SCFM) maximum
EOS/ESD Specifications:
The specifications shown below apply except on "Soft Ground Systems" which have
a 1 meg ohm current limiting resistance and a label placed on the power source front
panel referring to EN 100015-1.
Tip-To-Ground
Resistance: ...................... Less than 2 ohms.
AC Leakage: .................... Less than 2 millivolts RMS from 50Hz to 10MHz.
Transient Level: ............... Less than 500mV peak, out to 100MHz.
General Information
204°C to 482°C (400°F to 900°F) nominal.
±1.1°C (±2°F) at idle from set tip temp.
NOTE
PPM 100 Process Monitor.