KaVo ESTETICA E30 Short Instructions For Use page 24

Hide thumbs Also See for ESTETICA E30:
Table of Contents

Advertisement

Hygiene
Note
This allows the over-pressure to be released slowly from the water bottle. If the pres‐
sure supply to the equipment is not switched of, a permanent venting sound is audi‐
ble.
Note
The water in the water bottle needs to be changed at least one daily. In order to en‐
sure the efficiency of the measures, KaVo urgently recommends reprocessing the
water with KaVo OXYGENAL 6.
▶ For cleaning, rinse the water bottle with water at less than 50 °C.
Filling the water bottle
▶ Wipe up any spilled OXYGENAL immediately in order to prevent discolouration of the surfaces.
▶ Remove the protective hose from the tube of the KaVo OXYGENAL 6 dosing facility.
▶ Guide the tube through the orifice of the coarse sieve of the KaVo OXYGENAL 6 bottle.
▶ Screw the KaVo OXYGENAL 6 dosing facility tightly to the KaVo OXYGENAL 6 bottle.
▶ Rotate the orifice in the lid to a filling level of 1.5 litres.
Lid snaps-in in the filling position.
▶ Place the KaVo OXYGENAL 6 bottle on a level surface.
▶ Squeeze the KaVo OXYGENAL 6 bottle until the disinfection liquid rises above the red marker.
In response to over-filling, the excess quantity above the tick mark is suctioned off automatically
such that exact dosing is ensured.
22