KaVo ESTETICA E30 Short Instructions For Use page 18

Hide thumbs Also See for ESTETICA E30:
Table of Contents

Advertisement

Setting the handpieces
Setting options
▶ Press the "S4" key to select the ENDO options menu.
S1
S2
Parameters
Transmission factor
Parameter memory
Torque mode
Direction of motor rotation
Speed
Torque
The following parameters can be changed:
▪ Speed
▪ Torque
▪ Torque mode
Speed
The speed can be adjusted as follows:
▪ in steps of ten:
in the range from 100 rpm to 500 rpm
▪ in steps of fifty:
in the range from 500 rpm to 1,000 rpm
▪ in steps of 100:
in the range from 1,000 rpm to 6,000 rpm
Torque
1:1 transmission ratio
The torque can be changed as follows:
▪ in steps of 0.05 Ncm:
in the range from 0.15 Ncm to 2.5 Ncm
▪ in steps of 2%:
in the range from 1 % to 100 %
3:1 transmission ratio
S3
S4
Values
1:1, 3:1
P1,P2,P3,P4,P5,P6
TQ-Ctrl, Autoreverse, AutoRev/Forw.
R, L
100 to 6,000 rpm
Transmission ratio 1:1: 0.15 to 2.50 Ncm
Transmission ratio 3:1: 0.4 to 8.0 Ncm