  • Page 1 Instructions for use Spare cartridge for EXPERTtorque mini - 1.007.9458, 1.007.9457 Always be on the safe side.
  • Page 2 Distributed by: Manufacturer: KaVo Dental GmbH Kaltenbach & Voigt GmbH Bismarckring 39 Bismarckring 39 D-88400 Biberach D-88400 Biberach Phone +49 (0) 7351 56-0 www.kavo.com Fax +49 (0) 7351 56-1488...
  • Page 3: Table Of Contents

    Instructions for use Spare cartridge for EXPERTtorque mini - 1.007.9458, 1.007.9457 Table of contents Table of contents Safety ................................ 2 Purpose - proper use ........................2 Safety instructions ........................... 2 Product description ............................ 3 Technical data ..........................3 Exchanging the spare turbine ........................4 First use ..............................
  • Page 4: Safety

    ▶ Stop work and seek service support. Note KaVo accepts no liability for damage arising from repairs that have been carried out inappropriately. The following persons are authorised to exchange the spare turbine: ▪ Technicians at KaVo branches throughout the world ▪...
  • Page 5: Product Description

    Instructions for use Spare cartridge for EXPERTtorque mini - 1.007.9458, 1.007.9457 2 Product description | 2.1 Technical data 2 Product description Wrench, optional Used for the EXPERTtorque mini turbine(s) ▪ Spare turbine with wrench Mat. no. 1.007.9458 ▪ Spare turbine without wrench Mat. no. 1.007.9457 2.1 Technical data...
  • Page 6: Exchanging The Spare Turbine

    Instructions for use Spare cartridge for EXPERTtorque mini - 1.007.9458, 1.007.9457 3 Exchanging the spare turbine 3 Exchanging the spare turbine CAUTION Residues in the head or lid of the replacement turbine. Malfunctions. ▶ Remove any residues of a defective replacement turbine, such as washers, spring washers or other parts, from the head housing or lid.
  • Page 7 Functional problem. ▶ Make sure that the O-rings are positioned correctly in the recesses. ▶ Replace both O-rings. ▶ Spray O-rings with KaVo Spray and insert them in the head housing and cov‐ Insert the replacement turbine DANGER Not maintaining a torque of 100 Ncm when tightening the cover.
  • Page 8: First Use

    Instructions for use Spare cartridge for EXPERTtorque mini - 1.007.9458, 1.007.9457 4 First use 4 First use CAUTION Compressed air connection on devices. Dirty and moist compressed air causes premature bearing wear. ▶ Ensure that the cooling air is dry, clean and uncontaminated in accordance with EN ISO 7494-2.
  • Page 9: Function Test

    Instructions for use Spare cartridge for EXPERTtorque mini - 1.007.9458, 1.007.9457 5 Function test 5 Function test ▶ Insert a tool (cutter or grinder) and turn it radially by hand in different direc‐ tions, applying slight thumb pressure, to precisely align the O-rings.
  • Page 10: Tools And Consumables

    Instructions for use Spare cartridge for EXPERTtorque mini - 1.007.9458, 1.007.9457 6 Tools and consumables 6 Tools and consumables Obtainable from the dentalmed. specialist supplier Material summary Mat. No. Wrench 1.006.3384 O-rings 0.200.6172 Spring washer (larger outer diameter, shiny) 1.003.5900 Spring washer (smaller outer diameter, coloured 1.003.7576...