Information - Proximity To Certain Areas; Information On Use - Otto Bock C-Brace Instructions For Use Manual

Table of Contents

Advertisement

Available languages

Available languages

CAUTION
Using the orthosis with incorrect heel height of the shoe
In certain cases, this can lead to falling caused by:
► Altered swing phase triggering behaviour due to incorrect heel height.
→ Only use the orthosis with shoes that have the heel height specified at the initial fitting.
4.2.6 Information – proximity to certain areas
CAUTION
Proximity to sources of strong magnetic or electrical interference
This may lead to falling caused by:
► Malfunctioning of the orthosis.
→ Avoid proximity to sources of strong magnetic or electrical interference (such as high-tension power lines,
transmitters, transformers, computer tomography,...).
CAUTION
Exposure to high temperatures
This may lead to falling caused by:
► Malfunctioning of the orthosis.
► Breakage of load-bearing components.
→ Avoid exposure to extremely high temperatures (allowable temperature range see Page 77).

4.2.7 Information on Use

CAUTION
Improper mode switching
In certain situations, this can lead to falling caused by:
► Unintentional mode selection resulting in altered damping behaviour of the orthosis (e.g. under flexion load in
cycling mode).
→ The patient must stand securely during all switching processes.
→ Verify the desired functionality after switching (observe acoustic feedback signals).
→ Switch back to 1st mode when the activities in 2nd mode are concluded.
→ Take weight off the orthosis and correct switching if required.
CAUTION
Risk of pinching in the joint mechanism
This can lead to injuries caused by:
► Pinching the fingers or the skin when flexing the orthosis.
→ Do not reach into the joint mechanism while sitting down or flexing the orthosis.