Hold/Track Setup - Teledyne 6600 Operating Instructions Manual

Oil in water analyzer
Table of Contents

Advertisement

Oil in Water Analyzer
Every time the power is cycled, the homogenizer defaults to ON. So if
homogenizer was off and the power is cycled, the homogenizer will turn on.
Warning: Do not operate the "ultrasonic homogenizer" in the
3.3.9
The 6600 has ability to disable the analog outputs and freeze the display
while undergoing a scheduled or remote calibration. The 6600 will track
changes in the concentration if calibration is started through the front
panel. To setup this feature, the operator must:
1.
2.
Second System screen appears:
3.
HOLD by toggling Enter switch. By selecting the TRACK option, the analog
outputs are enabled and with the display will track the concentration changes
while the instrument is undergoing scheduled or remote calibration (either zero or
span). By selecting the HOLD option, the analog outputs and display are
disabled and will not track the concentration changes while the instrument is
undergoing scheduled or remote calibration (either zero or span). In the HOLD
option, the analog outputs and display will freeze on the last reading before
entering calibration.
The analog outputs are both 0 to 1 volt outputs and both 4 to 20 mA
outputs.
instrument for more than one (1) minute without a
liquid sample properly flowing through the homoge-
nizer.

Hold/Track Setup

Enter the System menu:
DIG_FILT
PWD LOGOUT
Using the UP/DOWN switch, select MORE and Enter. The
AUTOCAL HMGNZR TRACK
CAL-HOLDER-TIMER
AUTOCAL HMGNZR HOLD
CAL-HOLD-TIMER
The option on the right of the first row can be set to TRACK or
Teledyne Analytical Instruments
Operation 3
SELF-TEST
MORE
MORE
or
MORE
Part I
3-13