Teledyne 6600 Operating Instructions Manual page 138

Oil in water analyzer
Table of Contents

Advertisement

Part III
Guide for Oil in Water Analyzer Systems:
Please refer to your particular piping, outline, and
wiring drawings of your supplied system in the addendum
portion of this manual.
General requirements for the oil in water systems are'
Notes:- Sample must be representative and single
phase with no high particulates (<40 NTU) nor
slugs of oil present that could plug inlet lines.
Plugging may require flushing with utility water
and/or disassembly of components for mainte-
nance or cleaning.
- Sampling time delays must fit the process con-
trol requirements
A
Site requirements:
1
Protection from the elements: We recommend the
customer provide protection on the Enclosure
(especially on an offshore platform) concerning
the following: i.e., direct sunlight, wind sheltering
with suitable mechanical floor and wall supports
if possible.
2
Protection (shock mounts) from large vibrations
such as pumps and valve operations.
3
Always continuous and free flowing with proper
high enough sample inlet gravity fed lines for
prevention of cavitation and easy self-priming of
1-48 Part III
Commissioning and Start-up
Teledyne Analytical Instruments
Model 6600

Advertisement

Table of Contents
loading

Table of Contents