Oil in Water
Interface Panel for input/output functions related to remote
communications.
Reference drawings
3
4
See section for the Alarms Function
a)
2.2, Part I under electrical connections
c
DCS connections
1
Please refer to Section 1.5 Control
Unit
2
See Alarm Relays, section 2.2
3
See Digital Remote Cal Inputs for Zero
and Span; section 2.2
a
Outline drawing
-physical locations and sizes of electrical and
sample connections
b
Interconnect drawing (see addendum
drawings)
-indicates where electrical connections go
Sample Connections
a The customer connection to the oil in water analysis
system should enter in a downward position enabling
a constant available gravity fed non-freezing sample
with continuous flows capable of up to 5 liters per
minute at pressure provided.
b Assure there are no blockages nor fine filters at the
sample tap nor beyond up to the oil in water analysis
system.