4 Maintenance
Figure 4-1: Removing Fuse Block Cap and Fuse from Housing
2. Replace fuse by reversing process in step 1.
4.2

System Self Diagnostic Test

1. Press the System button to enter the system mode.
2. Use the < > arrow keys to move to More, and press Enter.
3. Use the < > arrow keys to move to Self-Test, and press Enter.
The following failure codes apply:
Table 5-1: Self Test Failure Codes
4-2: Part I
3.0 A MAX
Power
0
OK
1
5 V Failure
2
15 V Failures
3
Both Failed
Analog
0
OK
1
DAC A (0–1 V Concentration)
2
DAC B (0–1 V Range ID)
3
Both Failed
Teledyne Analytical Instruments
Model 6600 Oil in Water Analyzer