Draining The System; Testing The System; Calibration; Calibration Fluids - Teledyne 6600 Operating Instructions Manual

Oil in water analyzer
Table of Contents

Advertisement

Oil in Water Analyzer

2.2.7 Draining the System

In liquid analysis systems, the system return must terminate back to the
process or a safe area as the sample may be poisonous or corrosive. Olso, the
return pressure must be always sufficiently low enough from the inlet pressure
to maintain proper response times within the system.
2.3
Before plugging the instrument into the power source:
NOTE: Special designed systems may require checks under vacuum
Warning:
Power up the system, and test it by performing the following
operations:
1. Repeat the Self-Diagnostic Test, section 3.3.4, part I
2.4
Zero and span fluids must be made by the chemistry lab or certified zero
and span fluids bought from a supplier. The zero fluid must be the major
component of the sample, free from the component of interest.

Testing the System

Check the integrity and accuracy of the fluid connections. Make
sure there are no leaks.
Check the integrity and accuracy of the electrical connections.
Make sure there are no exposed conductors
Check that sample pressure is between 3 and 40 psig, according
to the requirements of your process.
or high pressure (consult manual addendum). Consult
commissioning start-up section in the manual addendum.
Do not operate the "ultrasonic homogenizer" in the
instrument for more than one (1) minute without a
liquid sample properly flowing through the homog-
enizer.

Calibration

2.4.1 Calibration Fluids

Teledyne Analytical Instruments
Part II: Analysis Unit
Part II: 2-3

Advertisement

Table of Contents
loading

Table of Contents