A
LARMS
Service Code
Alarm Name/Mitigation/Resolution
2090
PEEP Not Met/Circuit Leak
P
drops below the PEEP setting by 2 cm H
aw
be caused by a leak in the breathing circuit, exhalation valve, or patient airway..
Mitigation/Info: Check Patient Connection, Check Circuit For Loose Hose/Tube, Check Exhalation
Valve, Check Tube Placement/Cuff
**Replace Circuit**
2095
Insufficient Flow
The pressure target is not reached during the inspiratory period during pressure targeted
ventilation. Typically, this can occur when the Rise Time is set too low for the patient and their
respiratory mechanics.
Mitigation/Info: Pressure Target Not Met, Decrease Rise Time, Press/Hold BPM Button
**Ventilate With Volume Target**
2170
Spontaneous Breath-PIP High
The P
aw
Mitigation/Info: Pressure Exceeds Limit Setting, Check Circuit for Kinked Hose/Tube, Check for
Airway Obstruction, Suction Airway If Necessary, PIP High Limit Set Too Low?
**Manually Ventilate Patient**
2171
Spontaneous Breath-PIP Low
The P
aw
Mitigation/Info: Check Circuit for Loose Hose/Tube, Check Exhalation Valve, Check Tube
Placement/Cuff, Is Low Limit Set Correctly?
**Manually Ventilate Patient**
2172
Spontaneous Breath-V
The High V
Mitigation/Info: Check Circuit for Loose Hose/Tube, Check Exhalation Valve, Check Tube
Placement/Cuff, Is High Limit Set Correctly?
**Manually Ventilate Patient**
2173
Spontaneous Breath-V
The Low V
Mitigation/Info: Check Circuit for Kinked Hose/Tube, Check For Airway Obstruction, Suction Airway
If Necessary, Is Low Limit Set Correctly?
**Manually Ventilate Patient**
2300
Pulse Ox Module Failed
The pulse oximeter module fails while in use. There is no operator intervention. When the alarm is
active, "-- --" will display in the HR and SpO2 windows.
Mitigation/Info: Internal Failure, SpO2/HR Not Available, Turn Off Pulse Ox
**Replace/Service Ventilator**
2301
Internal Communication Failed
The communication between the pulse oximeter and the unit fails.
Mitigation/Info: Pulse Ox Module Failure, SpO2/HR Not Available, Turn Off Pulse Ox
**Replace/Service Ventilator**
4-16
exceeds the High PIP Limit setting during 2 consecutive spontaneous breaths.
exceeds the Low PIP Limit setting during 2 consecutive spontaneous breaths.
High
T
Limit is exceeded during 2 consecutive spontaneous breaths.
T
Low
T
Limit is not achieved during 2 consecutive spontaneous breaths.
T
www.zoll.com
O during the expiratory phase of the breath. This can
2
906-0731-01 Rev. J